Command line¶
AnyWave can be driven from a terminal. The command line lets you:
- open a data file (or a BIDS folder) in the interface,
- run a processing plugin without any window (ICA, h2/r2, spectral analysis, export...), file after file, from a script,
- convert a recording to BIDS,
- start AnyWave in the debug mode used by MATLAB and Python plugins.
The options of each processing plugin are described in Plugins on the command line.
Starting AnyWave from a terminal¶
Windows: add the folder of the AnyWave application to the system PATH, or give the full path of the application in your commands.
Linux: the Debian package (and a build from the sources) creates a link in /usr/bin/anywave: the anywave command is always found.
macOS: the installation package creates the link /usr/local/bin/anywave, so anywave works in the Terminal. The program itself is
/Applications/AnyWave.app/Contents/MacOS/AnyWave.
anywave # opens the interface
anywave recording.edf # opens the interface on a data file
anywave /data/my_bids_dataset # a folder is opened as a BIDS dataset
anywave --version # AnyWave version, build date and Qt version (also: -v)
anywave --help # also: -h
Older versions of --help
--help lists the options of this page. AnyWave 2.16.6 and earlier printed spellings that the program never accepted (--input, --args,
--plugin-debug, --server-port, --log-level, --output-log and the short forms -r, -i, -o, -l): they fail with Unknown option.
Use the names below: --input_file, --plugin_debug, --server_port, --log_dir...
Batch mode and interface mode¶
| command | what happens |
|---|---|
anywave --run <plugin> ... |
batch: no window. AnyWave runs the plugin, writes its results and exits |
anywave --toBIDS ... |
batch: converts a recording to BIDS, then exits (see below) |
anywave [file or folder] |
the interface opens (with the file, or the BIDS dataset) |
anywave --plugin_debug --server_port <port> [file] |
the interface opens and listens on <port> for a plugin being debugged |
In batch mode AnyWave returns an exit code (see Exit code).
Options¶
An option and its value are separated by a space: --hp 1. Paths with spaces must be quoted.
Option names are lowercase, with underscores, and are case sensitive. The values of --output_format and --create_montage, and the names of
plugins, are not.
Input¶
| option | value | description |
|---|---|---|
--input_file |
file path | the data file to process. Mandatory for every plugin that reads data |
--run |
plugin name, JSON file or JSON string | the processing plugin to run, see Running a plugin |
--marker_file |
file path | a .mrk file to use instead of the one next to the data |
--montage_file |
file path | a .mtg file to apply instead of the default montage |
--create_montage |
bipolar_ieeg, monopolar or none |
build the montage on the fly, see Channels |
--pick_channels |
list of labels | use only these channels, see Channels |
--modality |
eeg, meg, seeg, ieeg... |
the type of channels processed by the plugins that need one (ICA, spectral analysis...). --bids_modality is accepted too, and gives the same result when --modality is not given |
--skip_bad_channels |
true, false, yes, no (default true) |
false also processes the channels marked as bad |
--use_markers |
labels separated by commas | process only the data covered by markers with these labels |
--skip_markers |
labels separated by commas | process the data except what is covered by markers with these labels |
--hp, --lp, --notch |
frequency in Hz | high pass, low pass and notch filters applied to the input channels |
Output¶
| option | value | description |
|---|---|---|
--output_dir |
folder | where to write the results. It is created when it does not exist. Default: the folder of the data file |
--output_file |
name | name of the result file (without extension) |
--output_prefix, --output_suffix |
text | added to the name of the result file. Used by h2/r2, ICA and spectral analysis; ignored by the other plugins |
--output_format |
vhdr, edf, ades, matlab, fiff |
file format of the result of an export |
--output_format only accepts the formats above, and only when the writer of the format is installed. matlab needs the MATLAB writer, which is not part
of every AnyWave package: when a format has no writer, AnyWave stops before running anything, with an error and the exit code 255. A job file
that names such a format ("output_format": "matlab") is stopped by the export plugin, which logs Plugin matlab format not found, writes nothing and returns a non zero exit code too.
Log¶
| option | value | description |
|---|---|---|
--log_dir |
folder | the folder where AnyWave writes its log files (default: the AnyWave/Log folder of your home or documents folder) |
--log_level |
debug, info, warning, error |
drops the messages below this level, in the console and in the log files. Without it every message is kept. Another value is refused (exit code 255) |
Interface options¶
| option | value | description |
|---|---|---|
--plugin_debug |
listen for a plugin being debugged. Requires --server_port. See MATLAB debug and Python debug |
|
--server_port |
TCP port | port used by --plugin_debug |
Running a plugin¶
--run takes one of three things:
The name of the plugin, as it is called in AnyWave. Case does not matter (--run ICA and --run ica are the same plugin);
quote names that contain spaces.
anywave --run h2 --input_file recording.edf --time_window 4 --step 1 --max_lag 0.1
anywave --run "EEG Artifact Detector" --input_file recording.edf
A JSON file that holds the name of the plugin (key plugin) and any option, with the same names as on the command line:
{
"plugin": "export",
"input_file": "recording.edf",
"output_format": "vhdr",
"output_dir": "converted",
"hp": 1,
"lp": 40
}
anywave --run job.json
A JSON string, with the same content as the file. It is the most compact form in a script (in a Windows shell, prefer a file: quotes inside a JSON string must be escaped, and the escaping rules depend on the shell).
anywave --run '{"plugin": "export", "output_format": "edf", "output_dir": "converted"}' --input_file recording.edf
When an option is given both on the command line and in the JSON, the command line wins.
The Batch tool of the interface works this way: for each file it writes a job_NNN.json file and runs
anywave --run job_NNN.json --log_dir <log folder>. You can open one of these files to see what a job looks like, and run it again from a terminal.
Options of the plugins¶
Besides the options above, every plugin declares its own options. They come in three kinds:
| kind | on the command line | example |
|---|---|---|
| parameter | --name value |
--time_window 4 |
| switch | --name alone |
--save_comp_traces |
| flag | --name true or --name false (yes and no are accepted) |
--several_result_files true |
A flag needs its value
--several_result_files followed directly by another option takes that option as its value, and the other option is lost.
Always write true or false after a flag.
They are listed, plugin by plugin, in Plugins on the command line. An option that no plugin declares is refused:
Unknown option 'name' and exit code 255.
What AnyWave does before running a plugin¶
Data, markers and montage files¶
When it opens the data file, AnyWave looks next to it for a montage file and a marker file, which have the name of the data file followed by
.mtg and .mrk (recording.edf.mtg), and for the list of bad channels (recording.edf.bad).
Those files are used when they exist. --montage_file and --marker_file replace them.
Without a .mrk file there are no markers, except those already stored in the data file by the acquisition system.
Channels¶
The channels given to the plugin are chosen in this order:
--pick_channels: only the listed channels, in the order of the list (this ignores--montage_fileand--create_montage). A channel is writtenFp1; a bipolar derivation is written with a dash:Fp1-Czis the signal ofFp1minus the signal ofCz.--create_montage:bipolar_ieegbuilds the bipolar montage of the SEEG electrodes (A1-A2,A2-A3...). It stops with an error when the file has no SEEG channels;monopolarandnoneuse the channels as recorded.
--montage_file: the channels of that montage.- Otherwise the default montage: the one of the
.mtgfile when there is one, the channels as recorded when there is none.
anywave --run h2 --input_file recording.edf --pick_channels "A1,A2,A3" --time_window 4 --step 1 --max_lag 0.1
Runs h2 with the three channels A1, A2 and A3 only. --pick_channels "Fp1-Cz,Cz-O1" gives two derivations, whatever the format of the data file.
Bad channels are left out (the channels of the .bad file, and those the data file marks as bad). --skip_bad_channels false puts them back; they are
then added after the other channels.
Markers and filters¶
--use_markers and --skip_markers select the parts of the file to process. With the markers good (2 s to 5 s) and bad (6 s to 8 s)
in a file of 10 s:
| options | processed data |
|---|---|
--use_markers good |
3 s: from 2 s to 5 s |
--use_markers good,bad |
5 s: both segments |
--skip_markers bad |
8 s: all but 6 s to 8 s |
--hp, --lp and --notch filter the channels before the plugin receives them, whatever the plugin.
Results¶
Results are written in --output_dir; without it, next to the data file. A plugin that writes a single file
names it after --output_file, or after the data file when there is no such option.
Exit code¶
In batch mode the exit code tells whether the run could be done:
| code | meaning |
|---|---|
0 |
the plugin ran to the end |
non zero (255 on Linux and macOS, -1 on Windows) |
an option is wrong or unknown, the plugin does not exist, the input file is missing or cannot be read, the plugin could not start or stopped on an error (a mandatory option of the plugin is missing, no writer for the format, the result file cannot be written...), or a --toBIDS conversion failed |
A script can use it to find the files that failed:
for f in /data/study/*.edf; do
anywave --run export --input_file "$f" --output_format vhdr --output_dir /data/converted \
|| echo "failed: $f"
done
A plugin that stops on an error marks the run as failed: for instance the spectral analysis without --overlap (Missing --overlap argument in the log), trigger_writer
without --meg_file, or an export whose writer is missing. AnyWave 2.16.6 and earlier returned 0 in those cases. A plugin written
outside AnyWave (MATLAB, Python) or another plugin may still report a problem in the log and return 0: keep a log with --log_dir for the runs you cannot check by hand.
Windows scripts
If your shell goes on without waiting for AnyWave, start it with start /wait in a cmd window or Start-Process -Wait in PowerShell.
Convert to BIDS¶
--toBIDS converts one recording into a BIDS file set (data file, _channels.tsv and the .json file of the modality), then exits.
anywave --toBIDS --input_file recording.vhdr --bids_modality eeg --bids_sub 01 --bids_task rest \
--bids_ses pre --bids_run 1 --bids_format edf --output_dir /data/bids/sub-01/ses-pre/eeg
creates, in /data/bids/sub-01/ses-pre/eeg:
sub-01_ses-pre_task-rest_run-1_eeg.edf
sub-01_ses-pre_task-rest_run-1_channels.tsv
sub-01_ses-pre_task-rest_run-1_eeg.json
| option | value | description |
|---|---|---|
--bids_modality |
eeg, ieeg, meg or anat |
mandatory |
--bids_sub |
subject | mandatory. 01 gives sub-01 |
--bids_task |
task | mandatory, except for anat |
--bids_ses |
session | optional |
--bids_run |
run | optional for EEG and iEEG. Mandatory for a 4D/BTi MEG recording |
--bids_acq |
acquisition | optional |
--bids_proc |
processing | optional |
--bids_format |
edf or vhdr (default) |
format of the converted EEG or iEEG data |
--bids_output |
all (default) or sidecars |
sidecars writes only the _channels.tsv and .json files (and the events), without converting the data file. EEG and iEEG only |
--bids_suffix |
T1w, T2w or CT |
mandatory for anat: the kind of image |
--output_dir |
folder | where the files are written. The folder must exist. Default: the folder of the input file |
- For EEG and iEEG the recording is given by
--input_file. - For MEG give the folder of the acquisition with
--input_dir(a 4D/BTi run, a CTF.ds, an Elekta/MEGIN folder).--output_diris then mandatory. For an Elekta/MEGIN recording--input_filecan name one.fiffile of the folder. - For an anatomical image (
--bids_modality anat) give a DICOM folder with--input_dir, or a NIfTI file (.nii,.nii.gz) with--input_file, together with--bids_suffixand--output_dir. The result issub-<id>[_ses-<s>][_acq-<a>][_run-<r>]_<suffix>.nii.gzfor a DICOM folder (a NIfTI file keeps its own extension), and its.json. - A missing mandatory option stops AnyWave with an error and the exit code 255. A conversion that fails afterwards (no reader for the file, a wrong
--bids_format, a file that cannot be read...) also returns the exit code 255, with the reason in the log. AnyWave 2.16.6 and earlier returned0in that case. Create the output folder before the conversion.
Default behaviour in short¶
--runwithout--input_fileonly works for plugins that do not read data.- No
--output_dir: results go next to the data file. - No
--output_formatwith the export plugin: AnyWave ADES (.adesand.dat). - No montage or marker option: the
.mtgand.mrkfiles next to the data are used when they exist. - Bad channels are skipped.
Example¶
Export a recording to EDF, filtered between 1 and 40 Hz, keeping only the segments marked good:
anywave --run export --input_file /data/subject1/recording.vhdr \
--marker_file /data/subject1/review.mrk --use_markers good \
--hp 1 --lp 40 --output_format edf --output_dir /data/converted